Noninvasive intracranial pressure assessment by optic nerve sheath diameter : automated measurements as an alternative to clinician-performed measurements

dc.contributor.authorNetteland, Dag Ferner
dc.contributor.authorAarhus, Mads
dc.contributor.authorSmistad, Erik
dc.contributor.authorSandset, Else Charlotte
dc.contributor.authorPadayachy, Llewellyn
dc.contributor.authorHelseth, Eirik
dc.contributor.authorBrekken, Reidar
dc.date.accessioned2024-06-20T12:17:34Z
dc.date.available2024-06-20T12:17:34Z
dc.date.issued2023-02-01
dc.descriptionDATA AVAILABILITY STATEMENT : The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.en_US
dc.description.abstractINTRODUCTION : Optic nerve sheath diameter (ONSD) has shown promise as a noninvasive parameter for estimating intracranial pressure (ICP). In this study, we evaluated a novel automated method of measuring the ONSD in transorbital ultrasound imaging. METHODS : From adult traumatic brain injury (TBI) patients with invasive ICP monitoring, bedside manual ONSD measurements and ultrasound videos of the optic nerve sheath complex were simultaneously acquired. Automatic ONSD measurements were obtained by the processing of the ultrasound videos by a novel software based on a machine learning approach for segmentation of the optic nerve sheath. Agreement between manual and automated measurements, as well as their correlation to invasive ICP, was evaluated. Furthermore, the ability to distinguish dichotomized ICP formanual and automaticmeasurements of ONSD was compared, both for ICP dichotomized at 20 mmHg and at the 50th percentile ( 14 mmHg). Finally, we performed an exploratory subgroup analysis based on the software’s judgment of optic nerve axis alignment to elucidate the reasons for variation in the agreement between automatic and manual measurements. RESULTS : A total of 43 ultrasound examinations were performed on 25 adult patients with TBI, resulting in 86 image sequences covering the right and left eyes. The median pairwise difference between automatically and manually measured ONSD was 0.06mm (IQR −0.44 to 0.38mm; p = 0.80). The manually measured ONSD showed a positive correlation with ICP, while automatically measured ONSD showed a trend toward, but not a statistically significant correlation with ICP. When examining the ability to distinguish dichotomized ICP, manual and automatic measurements performed with similar accuracy both for an ICP cuto at 20 mmHg (manual: AUC 0.74, 95% CI 0.58–0.88; automatic: AUC 0.83, 95% CI 0.66–0.93) and for an ICP cuto at 14 mmHg (manual: AUC 0.70, 95% CI 0.52–0.85; automatic: AUC 0.68, 95% CI 0.48–0.83). In the exploratory subgroup analysis, we found that the agreement between measurements was higher in the subgroup where the automatic software evaluated the optic nerve axis alignment as good as compared to intermediate/poor. CONCLUSION: The novel automatedmethod ofmeasuring theONSD on the ultrasound videos using segmentation of the optic nerve sheath showed a reasonable agreement with manual measurements and performed equally well in distinguishing high and low ICP.en_US
dc.description.departmentSurgeryen_US
dc.description.librarianam2024en_US
dc.description.sdgSDG-03:Good heatlh and well-beingen_US
dc.description.sponsorshipThe South- Eastern Norway Regional Health Authority.en_US
dc.description.urihttp://www.frontiersin.org/Neurologyen_US
dc.identifier.citationNetteland, D.F., Aarhus, M., Smistad, E., Sandset, E.C., Padayachy, L., Helseth, E. & Brekke (2023) Noninvasive intracranial pressure assessment by optic nerve sheath diameter: Automated measurements as an alternative to clinician-performed measurements. Frontiers in Neurology 14:1064492. DOI: 10.3389/fneur.2023.1064492.en_US
dc.identifier.issn1664-2295 (online)
dc.identifier.other10.3389/fneur.2023.1064492
dc.identifier.urihttp://hdl.handle.net/2263/96569
dc.language.isoenen_US
dc.publisherFrontiers Mediaen_US
dc.rights© 2023 Netteland, Aarhus, Smistad, Sandset, Padayachy, Helseth and Brekken. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY).en_US
dc.subjectUltrasounden_US
dc.subjectAutomated measurementsen_US
dc.subjectMachine learningen_US
dc.subjectOptic nerve sheath diameter (ONSD)en_US
dc.subjectIntracranial pressure (ICP)en_US
dc.subjectTraumatic brain injury (TBI)en_US
dc.subjectSDG-03: Good health and well-beingen_US
dc.titleNoninvasive intracranial pressure assessment by optic nerve sheath diameter : automated measurements as an alternative to clinician-performed measurementsen_US
dc.typeArticleen_US

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Netteland_Noninvasive_2023.pdf
Size:
1.3 MB
Format:
Adobe Portable Document Format
Description:
Article

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: